| ITEM NO DHS-100 DHS-225 DHS-500 DHS-010 DHS-013 |
Size of studio D*W*H 450*450*500 500*600*750 800*700*900 1000*1000*1000 1000*1000*1300 |
| Temperature range |
0℃ --100℃ |
| Humidity range |
30-98% R.H |
| Fluctuations/uniformity |
≤±0.5℃ /≤±2℃ |
| Humidity deviation |
+2, -3% R.H |
| Heating time |
0℃ --100℃ about 35min |
| Cooling time |
100℃ ---0℃ about 40min |
| Controller: imported Microcomputer Integrated temperature and humidity controller |
| Precision range set precision: temperature±0.1℃ instructed precision: temperature±1% R.H humidity: ±1% R.H |
| Temperature and humidity sensor |
platonic resistance PT100 /MV |
| Heating system: all independent systems, Ni-Cr alloy, heating heaters |
| Add humidity system: External insulate style, evaporation in stainless steel surface type Add humidity machine |
| Dehumidification system: evaporator coil contacts dew-point temperature Dehumidification way laminar |
| Water supply system: automatically control to add humidity and supply water |
| And can recover the waste water to cut down expense |
| Cooling system: all :closed one way /multi ways wind-cooled Compression refrigeration |
| (brand: “TAI KANG” from France) |
| Cycling system: high temperature resistance, low noisy air conditioning motor |
| multi-leaf centrifugal wind round |
| Material of outside box: carbon steel plate of good quality, which is deal with electrostatic phosphate powder-coating,SUS304 stainless steel plate with veins in the surface |
| Material of inside box:SUS304 stainless plate with mirror surface of good quality |
| Material of heat preservation |
polyurethane foaming glass fiber cotton |
| Heat insulation doorframe: double high, low temperature-resistance, anti-aging rubber sealing bar |
| Standard configuration |
glass vision window for lighting(1set) |
| Sample rack 2pcs 1 test holes (25, 50,100mm) |
| Safety protections: have memory in situation of Leakage, short-circuit |
| over-temperature of the electric machinery and lack of water |
| Also have memory when Over-current/voltage of compressor happens. |
| Power supply voltage:AC380V±10% 50±0.5HZ three-phase five-wire system |
| Temperature of the environment |
5℃ --+30℃ ≤85% R.H |
P.S 1 “DHS”is the pattern symbol of low temperature constant temperature and humidity test chamber. 2, All the data are taken in condition of 25℃(QT) and the working studio has no load 3, We can make nonstandard low temperature constant temperature and humidity test chamber according to the demand of the customer. |